A portable standalone wireless electric cellsubstrate impedance sensing (ECIS) system for assessing dynamic behavior of mammalian cells

dc.contributor.authorKasiviswanathan, Uvanesh
dc.contributor.authorPoddar, Suruchi
dc.contributor.authorKumar, Chandan
dc.contributor.authorJit, Satyabrata
dc.contributor.authorMahto, Sanjeev Kumar
dc.contributor.authorSharma, Neeraj
dc.date.accessioned2020-10-14T09:47:31Z
dc.date.available2020-10-14T09:47:31Z
dc.date.issued2020-12-01
dc.description.abstractIn this study, we have developed a wireless, portable, standalone, and simple electric cell-substrate impedance sensing (ECIS) system to analyze in-depth functional aspects of cellular functions on the surface of a co-planar metal electrode coated on conventional glass substrate using a low-cost circuitry and correlated it with an equivalent electrical circuit (EEC) model. Low-cost circuitry was used for studying the dynamic behavior of the mouse myoblast cells (C2C12) in a culture chamber. Further, the developed ECIS system was connected with 8-bit Arduino UNO microcontroller board for establishing a compact sized measuring unit, which can be placed inside a CO2 incubator to provide proper environmental condition for the biological cells during the entire measuring time. Integrating ZigBee RF module with the 8-bit Arduino UNO microcontroller board provides a wireless communication network. Theoretical calculation of the lumped-elemental electrical parameters associated with cell-electrolyte interface and metal-electrolyte interface was calculated. The calculation was performed by fitting the experimental impedance data to EEC model using least mean square method to determine the dynamic and vital functions of the mammalian cells such as proliferation (in real-time) with a change in intrinsic electrical parameters associated at any particular time point. Impedance measurements and the lumped-elemental electrical parameter were correlated with the respective microscopic images. The developed ECIS system was found to enable measuring of the extent of cellular proliferation over time. The compactness of the developed ECIS system integrated with the ZigBee RF module and the 8-bit Arduino UNO microcontroller board facilitates its utilization even when placed in the CO2 incubator for a prolonged time. © 2020, The Author(s).en_US
dc.description.sponsorshipMinistry of Human Resource Developmenten_US
dc.identifier.issn2093-3134
dc.identifier.urihttps://idr-sdlib.iitbhu.ac.in/handle/123456789/802
dc.language.isoen_USen_US
dc.publisherSpringeren_US
dc.relation.ispartofseriesJournal of Analytical Science and Technology;
dc.relation.ispartofseries;vol. 11 issue 1
dc.subjectCellular functionen_US
dc.subjectLow-cost wireless systemen_US
dc.subjectImpedance measurementsen_US
dc.subjectCurve fittingen_US
dc.subjectEEC modelen_US
dc.titleA portable standalone wireless electric cellsubstrate impedance sensing (ECIS) system for assessing dynamic behavior of mammalian cellsen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
A-portable-standalone-wireless-electric-cellsubstrate-impedance-sensing-ECIS-system-for-assessing-dynamic-behavior-of-mammalian-cells2020Journal-of-Analytical-Science-and-Technology.pdf
Size:
2.29 MB
Format:
Adobe Portable Document Format
Description:
RESEARCH ARTICLE

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: