Repository logo
Institutional Digital Repository
Shreenivas Deshpande Library, IIT (BHU), Varanasi

An experimental and numerical study on nodular gel phantom during cryotherapy

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

BACKGROUND: Cryogen spray cooling is an effective method to treat nodular basal cell carcinoma. OBJECTIVE: To study the effect of spraying distance (z = 27 mm, 18 mm and 9 mm) on a nodular gel phantom and to validate the numerical model. MATERIALS AND METHODS: Liquid nitrogen is sprayed on agarose gel phantom using 0.8 mm nozzle diameter. A two-dimensional Pennes equation with phase change is solved on the axisymmetric non-orthogonal grid. RESULTS: The lethal front is obtained at 4 mm from the gel surface, irrespective of spraying distance and the same is validated with numerical results. The maximum necrotic volume and lethal front achieved by -50°C and -25°C isotherms seem to have a lower dependence on the spraying distance as opposed to those corresponding to the 0ºC isotherm. In contrast, the initiation of these parameters is highly dependent on the nozzle to gel height. CONCLUSION: The numerical study presents good agreement with experimental data. The decrease in spraying distance leads to higher rate of maximum ablation volume during freezing, but final ablation volume remained approximately similar. © CryoLetters.

Description

Keywords

Citation

Collections

Endorsement

Review

Supplemented By

Referenced By