Repository logo
Institutional Digital Repository
Shreenivas Deshpande Library, IIT (BHU), Varanasi

Slot-Loaded Pentagon Microstrip Patch Antenna for Hyperthermia Application at 434 MHz

dc.contributor.authorKhan A.; Dubey S.K.; Singh A.K.
dc.date.accessioned2025-05-23T11:17:47Z
dc.description.abstractIn this paper, a slotted pentagon ringtone patch antenna is proposed for the hyperthermia application. The proposed antenna operates in the ISM of 434 MHz and the size of the proposed antenna is 120 × 120 × 1.6 mm3, making it a more suitable applicator for hyperthermia treatment. The antenna structure is considered in the combined single-layer human tissue model and simulated in Ansys HFSS 2020R1 software. The gain obtained for the proposed antenna is − 9.1 dBi. The specific absorption rate is achieved for the x-axis variation inside the phantom at different penetration depths, and these are all within the limits of the IEEE standards. © 2023, Metrology Society of India.
dc.identifier.doihttps://doi.org/10.1007/s12647-023-00668-4
dc.identifier.urihttp://172.23.0.11:4000/handle/123456789/7763
dc.relation.ispartofseriesMapan - Journal of Metrology Society of India
dc.titleSlot-Loaded Pentagon Microstrip Patch Antenna for Hyperthermia Application at 434 MHz